Do you export medical devices and plan MDSAP single audit? Many manufacturers are confused: why does MDSAP audit take 5 days for some companies but 7 days for ours? What criteria determine audit days?
The answer lies in the core MDSAP document AU P0008 《Audit Time Determination Procedure》(Version AU P0008.008, released on 2023-10-13).
This document is the mandatory basis for MDSAP-recognized Auditing Organizations (AOs) to calculate audit man-days and prepare audit plans. Its supporting form is MDSAP AU F0008.2 《Audit Duration Calculation Form (Audit Model 2017)》.
https://www.mdsap.global/documents/library/audit-time-determination-procedure
AOs cannot arbitrarily set audit days based on experience.
The same calculation model applies to initial certification, surveillance audit and recertification audit.
Adjustments shall be made according to actual business, headcount and product risk of the enterprise.
Calculation records shall be retained as supporting evidence for the audit plan.
Note:The calculated result is the base audit time. AOs may add extra audit time with justifiable reasons to achieve audit objectives.
Audits without Stage 1:Identify applicable tasks directly, apply the algorithm and adjustment factors to get total audit man-hours.
| Total Employees | Audit Time Reduction |
|---|---|
| 1–5 | -42% |
| 6–10 | -39% |
| 11–15 | -35% |
| 16–25 | -28% |
| 26–45 | -18% |
| 46–65 | -14% |
| 66–85 | -10% |
| 86–100 | -5% |
Important Reminder: Reduction is not unlimited. Comprehensive assessment shall be performed considering product complexity, multiple sites, multiple product lines, previous non-conformities and special processes (sterile, implantable devices, etc.). AOs shall document all adjustment justifications and shall not reduce audit time merely by applying the headcount discount.
The final audit man-days shall be output and filed as an attachment to the audit plan.