EU Updates MDR/IVDR Harmonised Standards! 9 New MDR + 1 New IVDR Standards Officially Published
Views:106
Time:2026-11-13 14:43:05


In November 2026, the European Commission published two key Implementing Decisions in the Official Journal of the European Union: (EU) 2026/2479 (for MDR) and (EU) 2026/2481 (for IVDR), which introduce and update a set of MDR/IVDR harmonised standards. Compliance with harmonised standards grants the presumption of conformity required by the legislation, serving as a core reference for Chinese medical device manufacturers applying for EU CE certification and compiling technical documentation. This round covers 9 MDR harmonised standards (including 2 superseding older versions) and 1 IVDR harmonised standard (superseding its predecessor). Transition periods are specified for certain superseded standards. Manufacturers shall proactively upgrade their documentation and test protocols in advance.


原
文
链
接


https://eur-lex.europa.eu/eli/dec_impl/2026/2481/oj

https://eur-lex.europa.eu/eli/dec_impl/2026/2479/oj

一、New / Updated Harmonised Standards



9 New/Updated Harmonised Standards under MDR:
 
Standard No.Standard English NameRemarks
EN ISO 11137-1:2026Sterilization of health care products — Radiation — Part 1: Requirements for development, validation and routine control of a sterilization process for medical devicesReplaces the old version; also included in the IVDR harmonised list. Presumption of conformity for the old version expires on 10 May 2028
EN ISO 14155:2026Clinical investigation of medical devices for human subjects — Good clinical practiceReplaces the old version. Clinical protocols and clinical reports shall be updated to this edition. The old version expires on 10 May 2028
EN ISO 4074:2026Natural rubber latex male condoms — Requirements and test methodsNewly harmonised
EN ISO 10993-6:2026Biological evaluation of medical devices — Part 6: Tests for local effects after implantationNewly harmonised
EN ISO 10993-7:2026Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residualsNewly harmonised
EN ISO 15883-6:2026Washer-disinfectors — Part 6: Requirements and tests for washer-disinfectors employing thermal disinfection for non-critical medical devices and medical equipmentNewly harmonised
EN IEC 60601-2-22:2022+A11:2026Medical electrical equipment — Part 2-22: Particular requirements for the basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipmentIncludes Amendment A11:2026
EN IEC 60601-2-57:2026+A11:2026Medical electrical equipment — Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for treatment, diagnosis, monitoring and cosmetic useIncludes Amendment A11:2026
EN ISO 80369-1:2026Small-bore connectors for liquids and gases in healthcare applications — Part 1: General requirementsNewly harmonised

1 New/Updated Harmonised Standard under IVDR:
Standard No.Standard English NameRemarks
EN ISO 11137-1:2026Sterilization of health care products — Radiation — Part 1: Requirements for development, validation and routine control of a sterilization process for medical devicesReplaces the old version. Presumption of conformity for the old version expires on 10 May 2028

Only 1 harmonised standard is updated for IVDR in this round, sharing the same radiation sterilization standard with MDR.
二、Key Transition Timelines for Enterprises



Effective date of new standards:10 November 2026 (on the date of publication in the EU Official Journal). Enterprises may adopt the new standards immediately to obtain presumption of conformity.
Withdrawal date of presumption of conformity for old EN ISO 11137-1 and EN ISO 14155:2028‑05‑10

Highlight:

Before 10 May 2028, the old versions retain harmonised status, and both old and new editions are valid in parallel. After 10 May 2028, reliance solely on the old standards will no longer grant presumption of conformity, and Notified Bodies may reject technical documentation. 

Enterprises with radiation-sterilised medical devices or ongoing clinical studies must include standard upgrade in their work plan.

三、Interpretation of This Update



The two Implementing Decisions (EU) 2026/2479 and (EU) 2026/2481 mark a major supplement to the pool of MDR and IVDR harmonised standards, rather than simple version replacement. They cover clinical research, biocompatibility, sterilization, active optoelectronic devices, connectors, and washer-disinfectors, addressing the long-standing pain point for many enterprises that "no harmonised standard is available and alternative conformity demonstration routes have to be adopted".

(一)Two replacement standards carry long-term core risks: EN ISO 14155:2026 & EN ISO 11137-1:2026
These two foundational general standards for clinical research and radiation sterilization are extremely broad in scope, covering both MDR and IVDR. The EU sets a transition period ending on 10 May 2028, which offers a buffer for existing certificates and ongoing clinical trials. It does not mean unlimited use of old versions during the transition. Notified Bodies may already require enterprises to assess the applicability of the new standards during new technical reviews, surveillance audits and change assessments. For newly launched clinical trials and newly developed sterilisation processes, adopting the new versions is the prudent approach. Notably, the new EN ISO 14155 revises requirements for clinical protocol design, investigator responsibilities, adverse event tracking and data traceability. Enterprises running clinical trials need to check clinical protocols, SOPs and investigator brochures for deviations.

(二)Complementary biocompatibility standards: EN ISO 10993-6 and EN ISO 10993-7 added to harmonised list
This update draws high attention from manufacturers of implantable and tissue-contacting devices. Previously, some parts of the 10993 series lacked MDR harmonised status. When performing biological evaluation, enterprises could not directly rely on harmonised standards to claim presumption of conformity, and had to prepare extra justification documentation and face stricter scrutiny from Notified Bodies. The inclusion of these two standards now provides a harmonised basis for local implant reaction testing and EO residual testing, substantially reducing the burden of evidence for biological evaluation reports. Please note: harmonisation does not waive biological risk assessment; a full justification must still be completed based on product contact type, material and contact duration.

(三)Specialised standards for active light source devices released: EN IEC 60601-2-22 & EN IEC 60601-2-57 (including A11)
Medical laser and intense light source devices for therapeutic/cosmetic use are high-priority categories under MDR and frequent targets for Notified Body audits. The newly harmonised dedicated standards include the A11 amendment, updating requirements for light source hazard classification, protection, labelling and instructions for use. Manufacturers of cosmetic laser and photon devices shall carefully verify whether existing test reports cover the supplementary clauses of A11, instead of only checking the main standard version.

(四)Harmonisation of EN ISO 80369-1 for small-bore connectors reduces compliance uncertainty for tubing devices
Small-bore connectors form the foundation for misconnection safety, covering infusion, respiratory, anaesthesia and many other categories. Harmonisation of this standard greatly facilitates risk file preparation and design verification for tubing and connector manufacturers.

(五)IVDR only incorporates EN ISO 11137-1 in this round — a single supplement instead of broad update
The overall pool of harmonised standards for IVDR remains limited. This release contains no IVDR-specific harmonised standards, only the shared radiation sterilisation standard. IVD manufacturers should not be overly optimistic. Apart from sterilisation, most IVD products still lack harmonised standards, leading to heavy conformity demonstration and Notified Body review pressure.
四、Practical Recommendations for Chinese Export Manufacturers



  • Sort out applicable standard lists: Cross-check the updated list to identify whether your products fall under the scope of radiation sterilisation, clinical GCP, biocompatibility (local implant effects, EO residuals), medical laser / non-laser light source equipment, small-bore tubing connectors, washer-disinfectors.
  • Roadmap for technical documentation update: For new projects and new registrations, directly adopt the 2026 new harmonised standards. For ongoing projects and existing CE certificates, use the transition window until 10 May 2028 to complete revision of test reports, risk analyses and technical files. Avoid batch rectification right before the deadline.
  • Reminder for IVD manufacturers: EN ISO 11137-1 applies to IVDR as well. IVD products adopting radiation sterilisation also need to follow the version upgrade.
五、Conclusion



The evolution of EU harmonised standards directly affects CE technical documentation, test protocols and certificate maintenance. Export manufacturers are advised to continuously track standard timelines and plan version migration in advance to mitigate compliance risks caused by last-minute concentrated rectification. Wiselink Group will keep updating regulatory developments for MDR and IVDR, supporting Chinese medical device manufacturers for smooth global market access.Please contact Wiselink for further inquiries.